This manual covers the operation, inspection and maintenance of Henan Yongxing Special Equipment Co., Ltd. pressure vessels: single-door, pass-through and vertical saturated-steam autoclaves, steam storage vessels, expansion tanks, air receivers and oxygen storage tanks. It is written for plant operators, maintenance fitters and the authorised inspector who signs the periodic inspection record. It covers pressure-vessel law essentials, foundation and seismic restraint, door interlock testing, gasket and safety-valve care, cycle validation, and the daily-to-annual maintenance schedule. Where this text and the registration certificate differ, the certificate governs.
1. Scope and model coverage
The family covers quick-opening autoclaves for saturated-steam sterilisation at 121 °C and 134 °C, with the static vessels that sit beside them in a steam plant. Chamber design pressure starts at 0.6 MPa, a wide margin over the 0.21 MPa gauge needed for a 134 °C cycle, and rises to 2.5 MPa for high-pressure receivers. Jackets are rated 1.0 MPa unless the nameplate states otherwise.
| Model family | Capacity range | Design pressure / temperature | Heating medium | Typical duty |
|---|---|---|---|---|
| YX-AU-S single-door autoclave | 0.6–12 m³ chamber | 0.6 / 1.0 MPa, 150 °C | Saturated steam | Batch sterilisation of instruments, media and glassware |
| YX-AU-P pass-through autoclave | 1.5–20 m³ chamber | 0.6 / 1.0 MPa, 150 °C | Saturated steam | Clean-side / dirty-side transfer in barrier facilities |
| YX-AU-V vertical autoclave | 0.1–1.5 m³ chamber | 0.6 MPa, 150 °C | Saturated steam or electric | Small laboratories, media preparation, waste kill cycles |
| YX-SSV steam storage vessel | 2–30 m³ | 1.0 / 1.6 MPa, 200 °C | Boiler steam | Peak-demand buffer, surge suppression, wet-steam knock-out |
| YX-EXT diaphragm expansion tank | 0.5 m³–10 m³ | 0.6 / 1.0 MPa, 120 °C | System water | Volume compensation in hot-water and jacket circuits |
| YX-AR air receiver | 1–20 m³ | 1.0 / 1.6 / 2.5 MPa, 80 °C | Compressed air | Compressor buffering, instrument air, door actuation reserve |
| YX-O2 oxygen storage tank | 5–50 m³ | 2.5 MPa, 80 °C | Gaseous oxygen | Oxygen header storage, vaporiser buffer, peak supply |
Excluded: site electrical works beyond the terminal box, the boiler and cooling plant, room ventilation, cleanroom pressure regimes, and the sterile barrier wall of a pass-through installation.
Every vessel carries a nameplate showing design pressure, design temperature, volume, medium, manufacturer, date and registration number. Register the vessel before use, keep the certificate on site, present it at each periodic inspection, and use only an authorised inspection body and trained operators. The certificate states the inspection interval and governs over this manual.
2. Site and utility requirements before delivery
Confirm every item below before delivery. A short foundation or a wet steam supply delays handover and cannot be fixed on site. Hold the room at 15–30 °C, frost-free and dry, with six air changes per hour.
Steam quality matters more than quantity. Carry a dedicated line from the boiler header and fit a separator and trap at the lowest point; wet steam carries non-condensable gas into the chamber and lengthens the cycle.
| Item | Requirement | Tolerance / notes |
|---|---|---|
| Foundation / plinth | Reinforced concrete, cube strength ≥ 25 MPa at 28 days, mass at least 1.5× vessel empty mass | Flat to ±3 mm over 3 m; plinth at least 100 mm above finished floor |
| Clearances | Door swing plus 1.2 m at the loading face; 0.8 m at the service face | Tube withdrawal space per GA drawing |
| Steam supply | Dry saturated, 0.4–0.6 MPa at the battery limit, sized for peak cycle | Separator and trap within 3 m of the inlet; 40-mesh strainer |
| Cooling water | 0.2–0.3 MPa at the jacket inlet, ≤ 30 °C, ≥ 2 m³/h per m³ of chamber | Chloride ≤ 25 mg/L for stainless chambers; strainer fitted |
| Compressed air | 0.5–0.7 MPa, dew point ≤ 2 °C, oil-free | Filter-regulator and lockable isolation at the vessel |
| Electrical supply | 400 V three-phase plus neutral and earth, dedicated isolator at the door | Disconnection within 0.4 s; bonding ≤ 0.1 Ω |
| Drainage | Blowdown pit or trench with air gap, capable of 90 °C discharge | Pipe slope ≥ 1:100 toward the pit; no lift in the drain line |
| Environment | 15–30 °C, ≤ 85% RH, no corrosive vapour | Oxygen tanks: ≥ 5 m from oil storage, no ignition source |
3. Installation sequence
- Survey the plinth before unloading. Check strength records, level to ±3 mm over 3 m and anchor bolt positions; stop if out of tolerance.
- Lift only at the marked lugs with certified slings and a rated spreader beam. Doors locked; never lift by the door frame, locking ring or a nozzle.
- Set the vessel on 25–50 mm grout pads and level to ±2 mm/m in both axes. Confirm the door face is square to the trolley rails.
- Fit anchor bolts and seismic brackets. Tighten to the drawing torque after the grout reaches 75% strength, then re-check level.
- Connect steam through an isolation valve, 40-mesh strainer and trap set. Support the pipework independently so no load reaches the nozzle.
- Pipe the safety valve outlet full bore to a safe point with no intervening valve; drill a 6 mm drain hole at the low point.
- Connect chamber and jacket drains to the pit through an air gap, falling at least 1:100 with no low pockets.
- Connect cooling water through a strainer and pressure-reducing valve; set the jacket inlet to 0.2–0.3 MPa and confirm return flow.
- Connect instrument air at 0.5–0.7 MPa through a filter-regulator with a lockable isolation valve.
- Complete the electrical termination and bond the vessel to the earth bar; verify ≤ 0.1 Ω and route signal cables clear of steam lines.
- Fit insulation only after the hydrostatic test, leaving the door flange, nameplate, safety valve and instruments clear.
- Hydrostatic test at 1.5× design pressure for 30 min: no leakage, no deformation, no drop beyond gauge error; then drain and dry.
- Prove every interlock in section 4, record the results, and hand over the registration documents before the first production cycle.
A loaded autoclave is top-heavy and its centre of gravity moves toward the door end. Use a spreader beam, keep personnel outside the swing radius, and never stand under a suspended vessel. A dropped vessel is a pressure-boundary defect: do not return it to service until the shell and door ring are inspected and the inspection body accepts it in writing.
Fix the vessel at the door end and let the far end slide on PTFE pads, or use slotted holes, so thermal growth of 1–2 mm/m does not load the shell. Seismic brackets are shimmed to 1 mm clearance and carry load only during an event. Never weld brackets or supports to a stressed shell without written approval.
4. Commissioning and handover
Hydrostatic test pressure is 1.5× the nameplate design pressure: 0.9 MPa for 0.6 MPa, 1.5 MPa for 1.0 MPa, 2.4 MPa for 1.6 MPa and 3.75 MPa for 2.5 MPa. Hold for 30 min after shell and water equalise in temperature, per GB/T 150 and the TSG 21 registration regime; boiler-side parts of a packaged supply follow GB/T 16508 and TSG G0001. Use water with chloride ≤ 25 mg/L, vent all high points, and never use the safety valve as a test vent.
Fill the jacket and chamber, vent until a steady stream runs, raise pressure in 25% steps with a 2 min hold at each, then drain and dry; standing water pits the gasket seat.
Set the steam supply so the chamber reaches 134 °C within the ramp time on the cycle card, normally 8–15 min for 2 m³. Check inlet steam quality with a conductivity meter or calorimeter; below 0.95 dryness means the separator or trap is not working. Compare chamber temperature with the saturation temperature for the measured pressure: a difference above 0.5 °C indicates air trapping.
| Interlock | Simulated condition | Expected result |
|---|---|---|
| Door open with pressure | Raise to 0.05 MPa; attempt to release the locking ring by hand and by control | Door stays locked; ring does not move; alarm sounds above 3 kPa |
| Door not locked, steam admitted | Leave the door ajar by 10 mm and call for a cycle | Steam valve stays closed; cycle refuses to start with a door fault |
| Pressure switch failure | Disconnect the switch signal during a hold phase | Cycle aborts, steam valve closes, vent opens, alarm sounds |
| Over-temperature | Force the sensor reading 3 °C above set point | Steam valve closes within 2 s; no re-admission until reset |
| Safety valve | Raise to set pressure with the door closed and chamber empty | Lifts at set pressure, reseats below 90% of set, accumulation ≤ 10% |
| Vacuum breaker | Cool a sealed chamber from 134 °C with the vent closed | Opens at −3 to −5 kPa; vacuum never below −10 kPa |
| Emergency stop | Press the E-stop during the sterilisation hold | Steam and air isolate, vent opens, cycle logged as aborted |
Validate with thermocouples in the coldest loaded position and with biological indicators of Geobacillus stearothermophilus at 106 spores per carrier. Three consecutive cycles must show F0 ≥ 15 min at 121 °C, a chamber spread within ±1 °C and no growth. Record the load pattern; validation applies only to that load.
Run a 4 h and then a 24 h trial at production load, logging ramp, hold temperature, jacket pressure, cooling time and condensate flow per cycle. Handover documents: registration certificate, nameplate rubbing, GA and pressure-boundary drawings, material certificates, weld and NDT reports, hydrostatic record, safety valve and calibration certificates, interlock record, validation report, spare gasket set and this manual.
5. Normal operation
Start-up from cold. Open the chamber drain and confirm it is clear. Open the steam isolation valve slowly and warm the jacket to 0.1 MPa over 10 min. Clean and dry the gasket seat, load below the marked load line, close and lock the door, confirm the lock indicator, then start the validated cycle.
Start-up from hot standby. Confirm jacket pressure and trap discharge. Vent the chamber for 60 s before loading, then close the drain. Do not open a hot door onto a cold load; thermal shock distorts the door ring and shortens gasket life.
Running log. Record each cycle: 0.11 MPa at 121 °C or 0.21 MPa at 134 °C, hold within ±1 °C, jacket 0.3–0.6 MPa, inlet steam 5–10 °C above saturation, cooling water ≤ 30 °C, and cycle time within 10% of the validated value. Rising temperature with falling pressure means non-condensable gas; a hold short of set point means a fouled trap or wet load.
Normal shutdown. Complete the cycle, vent to zero, confirm the gauge reads zero and the vent is open, then release the interlock and open the door. Close the steam isolation valve, drain the jacket if idle, and leave the door ajar.
Emergency shutdown. Press the E-stop, close the manual steam valve, and leave the chamber sealed until the gauge reads zero. Do not force the door. If pressure does not fall within 15 min, treat it as a blocked vent; never open a fitting on a pressurised shell.
Expansion tank. Set the nitrogen pre-charge to 0.1 MPa below the system static pressure before filling, and re-check it quarterly with the tank isolated and drained.
Lay-up over 30 days. Wet: fill with demineralised water at pH 10–11, add oxygen scavenger to 200 mg/L and circulate monthly. Dry: drain, dry with warm air, add 1 kg of silica gel per m³ through the inspection opening, and blank the connections. Leave the safety valve fitted and tagged, never gagged. Renew the gasket and re-prove the interlock before return to service.
6. Maintenance schedule
The schedule assumes daily production use. Increase frequency where steam quality is poor, the load is corrosive or the room is humid. Internal work requires a cold, vented, isolated and gas-free vessel under permit.
6.1 Daily (each shift)
| Task | What to check / acceptance | Tool or reference | Who |
|---|---|---|---|
| Door seal visual | No cuts or hardening; seal uniform in its groove; no steam plume at the door edge | Torch, cycle log | Operator |
| Zero-pressure interlock check | At zero gauge, door opens only with the vent open and lock indicator clear; no bypass key | Function check, SOP-04 | Operator |
| Pressure gauge reading | Zero before start; tracks the controller within 0.02 MPa | Gauge, HMI display | Operator |
| Steam trap | Intermittent discharge, not live steam; trap body hot both sides | Infrared thermometer | Operator |
| Chamber drain strainer | Blow down 5 s; flow free, water clear, no scale | Drain valve | Operator |
| Insulation surface | Cladding intact; surface ≤ 60 °C at 1 m height, ≤ 45 °C on door faces | Contact thermometer | Operator |
| Cycle record | Ramp, hold and F0 within 10% of validated; aborts logged with cause | Chart or HMI export | Operator |
6.2 Monthly
| Task | What to check / acceptance | Tool or reference | Who |
|---|---|---|---|
| Safety valve manual lift | Lift the lever at ≥ 75% of set; opens and reseats cleanly, no simmer or leak | Lever, set-pressure card | Maintenance fitter |
| Gauge comparison | Compare with a master gauge of 0.4 class; deviation ≤ 1.6% of full scale, otherwise replace | Master gauge, calibration tag | Instrument technician |
| Gasket and door ring | Compression even, no flattening beyond 25%; ring and groove clean, no wire drawing | Straight edge, 10× loupe | Maintenance fitter |
| Hinge and locking lubrication | Hinge pin and locking-ring teeth greased; door swings with one hand; no binding at the last 20° | High-temperature grease, OEM list | Maintenance fitter |
| Vacuum breaker | Remove, clean and test the disc; opens at −3 to −5 kPa, seals at zero | Vacuum test rig | Instrument technician |
| Interlock pressure switch | Switch changes state at 3 kPa rising and 2 kPa falling; repeatability within 0.5 kPa | Hand pump, multimeter | Instrument technician |
| Air receiver drain | Drain until clear; no oil carry-over; auto-drain timer at 4 h | Drain valve, timer log | Operator |
6.3 Quarterly / semi-annual
| Task | What to check / acceptance | Tool or reference | Who |
|---|---|---|---|
| Gasket replacement | Replace if hardness > 70 Shore A, compression set > 25% or age > 12 months; silicone to 200 °C, EPDM below 150 °C; torque in three passes at 50/80/100% of 45 N·m in a cross pattern | Durometer, torque wrench | Maintenance fitter |
| Door hinge and pin wear | Pin and bush clearance ≤ 0.2 mm; replace bush if the door drops more than 1 mm at the free edge | Vernier, dial gauge | Maintenance fitter |
| Chamber temperature mapping | Nine-point map within ±1 °C at 121 °C; coldest point on the cycle card | Calibrated thermocouples | Validation engineer |
| Safety valve pop test | Lift pressure within ±3% of set; reseat pressure ≥ 90% of set; no leakage at 90% of set | Test bench or steam rig | Authorised test body |
| Condensate drainage | Drain line fall ≥ 1:100; no sagging; trap strainer clean; blowdown pit free of scale | Level, visual | Maintenance fitter |
| Insulation and cladding | No damp patches, no gaps at flanges; surface temperature re-checked at three points | Thermal camera | Maintenance fitter |
6.4 Annual shutdown inspection
| Task | What to check / acceptance | Tool or reference | Who |
|---|---|---|---|
| Internal inspection | No cracks or pitting > 0.5 mm; welds sound; scale < 0.3 mm | Borescope, wet fluorescent MT | Authorised inspector |
| Hydrostatic test | 1.5× design, 30 min; no leak or deformation, drop within gauge error | Test pump, calibrated gauge | Authorised inspector |
| Wall thickness check | Readings at the marked grid; wall ≥ design plus corrosion allowance, and above retirement thickness | UT thickness gauge | NDT technician |
| Door locking ring NDT | Dye penetrant or magnetic particle on teeth and threads; no indication > 2 mm | PT / MT kit | NDT technician |
| Safety valve certification | Strip, clean, lap seat, re-set; certificate with lift and reseat pressures | Valve workshop | Authorised test body |
| Gauge calibration | Calibrate to 1.6 class; certificate within 12 months; red line at set pressure | Dead-weight tester | Instrument technician |
| Full interlock test | All seven section 4 tests repeated and signed; no bypass; E-stop functional | Test record | Maintenance engineer |
Every three years, extend the annual thickness check to a full survey: a marked grid at 200 mm pitch on the shell, heads, door ring and jacket, recorded against the previous survey. Loss above 0.1 mm/year triggers a review of water treatment; a wall near retirement thickness requires re-rating or replacement. The comprehensive inspection interval is that on the registration certificate, normally not more than three years for a quick-opening autoclave.
Keep cycle logs, interlock sheets, calibration certificates and inspection reports for at least three years, longer where the authority requires. Escalate any anomaly affecting the pressure boundary, interlock or validated cycle the same shift; never sign off a deviation as “monitor” without a written decision.
7. Troubleshooting
| Symptom | Likely cause | Check | Remedy |
|---|---|---|---|
| Steam leaks at the door during hold | Gasket worn, hardened or badly seated; door ring damaged | Seal groove, gasket hardness, ring surface | Replace gasket and torque in three passes; replace ring if grooved |
| Chamber will not reach 134 °C | Wet steam, blocked trap, air pocket, low steam pressure | Inlet pressure and temperature, trap discharge, NCG | Clear trap, vent chamber, raise header pressure, fit separator |
| Temperature and pressure disagree | Non-condensable gas in the chamber | Saturation table comparison; air-removal time | Extend venting, check vacuum breaker seal, service trap |
| Vacuum will not hold after cooling | Door seal leak, breaker stuck open, drain passing | Leak test at −50 kPa, breaker disc | Replace seal, clean breaker seat, reseat drain valve |
| Slow cooling | Fouled jacket, low water flow, air lock | Jacket inlet/outlet temperature, flow | Descale jacket, vent high point, clean strainer |
| Safety valve weeps at normal pressure | Dirt on seat, spring relaxation, low set pressure | Seat condition, set-pressure certificate | Clean and lap seat, re-set and certify; never exceed design pressure |
| Water hammer in steam line | Condensate pooling, trap failed closed, sagging pipe | Trap function, pipe fall, low-point drains | Replace trap internals, correct fall to 1:100, add drain pocket |
| Door interlock will not release | Residual pressure, switch stuck, vent blocked | Gauge reading, vent flow, switch state | Vent fully; replace switch; clear vent line. Never force the door |
| Pressure falls overnight on a receiver | Leaking fitting, failed check valve, open drain | Soap test on joints, drain position | Remake joints, replace check valve, close drain |
| Gauge reads high at zero pressure | Pointer shifted, blocked impulse line | Master gauge comparison, impulse line | Replace gauge, clear impulse line, recalibrate |
8. Spare parts and service
Hold these first-year spares on site. Items marked critical stop a cycle after a single failure.
- Door gasket set per chamber size, plus a spare for the pass-through transfer door (critical).
- Safety valve of the same set pressure and bore, with certificate (critical).
- Vacuum breaker assembly and two replacement discs.
- Interlock pressure switch and one door-position limit switch.
- Pressure gauge, 1.6 class, correct range, red set-pressure line.
- Steam trap internals kit and 40-mesh strainer element.
- Door hinge bush and pin set; locking-ring lubricant to the OEM list.
- Chamber and jacket thermocouples, plus one spare probe.
- Consumables: gasket sheet, O-rings, graphite packing, descaler, biological indicators and indicator tape.
Order by model, serial number, drawing number and item position on the drawing. Gaskets and safety valves are not interchangeable between sizes. State design pressure and medium when ordering a valve; an air receiver valve must never be fitted to a steam vessel.
Service contact: Henan Yongxing Special Equipment Co., Ltd. service desk, quoting the serial number from the nameplate. Remote support answers within one working day; a site visit for a pressure-boundary or interlock fault is scheduled within three working days. Annual inspection and hydrostatic testing can be booked with the third-party inspection body through the same desk. See the manual library for the current revision and the product range for model data.
9. Document control
| Rev | Date | Change |
|---|---|---|
| A | 2026-05-12 | First issue for the pressure-vessel family |
| B | 2026-11-25 | Added pass-through interlock tests, 3-year thickness survey and oxygen tank siting note |
This file is hash-verified. The SHA-256 digest of the published document is listed on the verification page; compare it with the digest of the file you hold before relying on any revision. If they differ, discard the copy and download the current file. Check the digest at https://www.yxboiler.com/manuals/verify/.
Related reading: capacity and sizing notes, steam and vessel calculators and energy benchmarks. For the autoclave range see autoclaves and steam sterilizers; for water treatment see softeners and feed pumps. Safety devices follow ISO 4126; receiver sizing follows US DOE guidance.
This is the generic family manual. The site-specific manual carries your serial number, registration data, validated cycle card and as-built drawing list.
Request the site-specific manual and quote the nameplate serial number.
Document control and download
Document
YX-MAN-PVS · revision 1.0 · issued 2026-11
Applies to the model families listed in section 1.
Download
Manual PDF · 229 KB
Print-ready A4, same content as this page.
Weekly maintenance log sheet · one page for the shift file.
SHA-256 ec708e2a7b2d859ca79cbf83e963ab13c7152afa8ae4f570a88d6fd18c802a7b
This fingerprint is calculated from the PDF file itself. Verify it on the verification page — the check runs in your browser, so an altered file cannot pass. A signed copy of the full manifest is published alongside it.
Key facts
- YXBOILER is the global brand and manufacturing site of Henan Yongxing Boiler Manufacturing Co., Ltd., an industrial boiler and pressure-vessel manufacturer founded in 1981 in Henan, China.
- Main products: steam boilers, hot-water boilers, thermal-oil heaters, biomass and coal-fired boilers, electric boilers, pressure vessels and autoclaves.
- Serving customers in worldwide with selection, drawings, export packaging, installation guidance and spare parts.
- Contact: WhatsApp +86 133 5377 2661 · [email protected].
Related on this site
Specifications, categories and further reading.
- About the company
- Contact & quotation
- Frequently asked questions
- Industrial boiler rangeSteam, hot-water, thermal-oil and biomass boilers.
YXBOILER network
The same manufacturer, documented site by site.
- YXBOILER product catalogueDetailed specifications and model pages for every boiler and pressure vessel.
- Yongxing Boilers — news & official updatesCorporate news, exhibitions, case studies and company announcements.
- YXBOILER IndonesiaBoiler supply, installation and service for Indonesian plants.
- Boiler MalaysiaIndustrial boiler supply and after-sales service in Malaysia.
- Boiler PhilippinesIndustrial boiler supply and after-sales service in the Philippines.
Standards & references
Independent sources used in our engineering work.
- ASME Boiler and Pressure Vessel CodeThe international reference code for boiler and pressure-vessel design and fabrication.
- ISO 9001 quality managementThe quality-management standard our manufacturing system is certified against.
- Spirax Sarco — learn about steamIndependent engineering reference for steam, condensate and heat-transfer practice.
- Boiler — WikipediaBackground on boiler types, fuels and thermodynamic principles.
YXBOILER — Global brand and manufacturing site. Last updated November 2026.